Measurement ranges |
AC Voltage: 100/200/500 V f.s. 3 ranges AC Current: 10 A to 100 A AC 4 ranges (with optional CLAMP ON SENSOR 9650) 10 A to 500 A AC 6 ranges (with optional CLAMP ON SENSOR 9651) 10 A to 1000 A AC 7 ranges (with optional CLAMP ON SENSOR 9668) |
Sampling period |
10 ms |
Frequency characteristic |
AC/DC voltage range: +0.5 dB to -3 dB from 30 Hz to 30 kHz. AC current range: Frequency characteristic is determind by the clamp sensor. |
Paper feed speed |
60 cm/hour to 2 cm/hour, 5 ranges |
Number of input channels |
1 channel AC voltage and 1 channel AC current (Simultaneous recording based on two channel alternating sampling, for commercial power lines, input is voltage isolated.) |
Accuracy |
Voltage: ±2 % f.s. (45 Hz to 66 Hz), Current: ±3.53 % f.s. (used with CLAMP ON SENSOR 9651 / option, AC 500 A range) |
Power supply |
100 to 240 V AC (50/60 Hz) or 9.5 to 14 V DC, 2 way |
Dimensions, mas |
250 mm(9.4 in)W × 122 mm(4.8 in)H × 93.5 mm(3.66 in)D, 1.2 kg (42.33 oz) |
Accessories |
Recording paper 9235 (1 roll), roll paper holder (2), Power cord (1), CONNECTION CORD 9257 (1), CARRYING CASE 9344 (1) |